Error Model and Parameter Identification for Serial-Parallel Cancer Curing M achines Using Ultrasonic

Fang Qiang1  Fan Liang-zhi2   

  1. 1.College of Mechanical Engineering,South China Univ.of Tech.,Guangzhou 510640,Guangdong,China;2.School of Mechanical Science and Engineering,Huazhong Univ.of Science and Tech., Wuhan 430074,Hubei,China
  • Received:2004-09-22 Online:2005-06-25 Published:2005-06-25
  • Contact: 方强(1974-),男,博士生,主要从事先进制造技术、机器人设计理论与工程应用方面的研究 E-mail:hgfq@ 163.com
  • About author:方强(1974-),男,博士生,主要从事先进制造技术、机器人设计理论与工程应用方面的研究

Abstract:

The actual position accuracy of the 5-DOF (Degree of Freedom)serial-parallel manipulator 0f a HIFU(High-Intensity Focused Ultrasonic)curing machine is analyzed on the basis of the actual error models of the mo-ving platform,the stability platform and the extension pole.as well as by means of the redundant restriction param-eter identification.With a 3-RPS serial-parallel manipulator as the research object.the model parameters of the
controllers of the curing machine are adjusted via the error parameter identification and position accuracv arialysis.and a higher position accuracy is achieved,thus meeting the positioning requirements for HIFU treatment more sat-isfactorily.The simulated results indicate that,during the practical measurement,about 50 test points are enough.with a maximum total eror of the focus less than 0.5 mm and a maximum measurement eror less than 0.2 mm,e-yen within 0.05 mm.The most remarkable merit of the proposed method is that no more extra configurati0n trans-ducers or sensors are needed during the parameter identification,which makes it possible for the manipulator to run on line in practical situations.

Key words: serial-parallel manipulator, error model, parameter identification, high-intensitv focused uhrasonic